Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Astro-Med PM-10: CAL ADJ /5520,34401 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 120 CONFIGURATION: Fluke 5520A CONFIGURATION: HP 34401A ============================================================================= # # Source: # Astro-Med PM-10 Calibration Procedure TP# 3163000X Rev D. # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 ASK- N P J F W 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Turn unit ON. 1.005 DISP 1.005 DISP Warm-up time: 30min - REQUIRED 1.005 DISP Temperature : 23 +/-2 degC 1.005 DISP Humidity : 55 +/-10% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD SET-UP PROCEDURE 1.008 DISP Insure that the PM-10 is not installed in a 1.008 DISP channel directly following a PM-08. 1.009 DISP Connect the 34401A to the UUT as follows: 1.009 DISP [32] 34401A Input HI to TP1 1.009 DISP [32] 34401A Input LO to GROUND (Pin 5 of U7) 1.010 DISP Note the position of DSW1 dip switches to determine 1.010 DISP the type of thermocouple the PM-10 is configured for. 1.010 DISP 1.010 DISP K = OFF ON OFF OFF 1.010 DISP J = OFF ON OFF ON 1.010 DISP E = OFF ON ON OFF 1.010 DISP T = OFF OFF OFF ON 1.011 MEM2 Enter thermocouple type (K, J, E, or T). 1.012 JMPL TYPE_K ZCMPI(MEM2, "K") 1.013 JMPL TYPE_J ZCMPI(MEM2, "J") 1.014 JMPL TYPE_E ZCMPI(MEM2, "E") 1.015 JMPL TYPE_T ZCMPI(MEM2, "T") 1.016 DISP Not a valid entry! 1.017 JMP 1.010 1.018 LABEL TYPE_K 1.019 HEAD TYPE K THERMOCOUPLE 1.020 DISP Connect the 5520A TC Terminals to the PM-10 input using 1.020 DISP TYPE K thermocouple wire and connectors. 1.021 RNG 500degC 1.022 5520 0degC _K TC S 2W 1.023 MESS Adjust OFFSET pot R69 for 0 +/-5mV. 1.024 34401 0mV S 2W 1.025 5520 500degC _K TC S 2W 1.026 MESS Adjust R21 for -5.0000 +/-0.0005V. 1.027 34401 -5.0000V S 2W 1.028 JMP 1.061 1.029 LABEL TYPE_J 1.030 HEAD TYPE J THERMOCOUPLE 1.031 DISP Connect the 5520A TC OUTPUT to the PM-10 input using 1.031 DISP TYPE J thermocouple wire and connectors. 1.032 RNG 500degC 1.033 5520 0degC _J TC S 2W 1.034 MESS Adjust OFFSET pot R69 for 0 +/-5mV. 1.035 34401 0mV S 2W 1.036 5520 500degC _J TC S 2W 1.037 MESS Adjust R21 for -5.0000 +/-0.0005V. 1.038 34401 -5.0000V S 2W 1.039 JMP 1.061 1.040 LABEL TYPE_E 1.041 HEAD TYPE E THERMOCOUPLE 1.042 DISP Connect the 5520A TC Terminals to the PM-10 input using 1.042 DISP TYPE E thermocouple wire and connectors. 1.043 RNG 200degC 1.044 5520 0degC _E TC S 2W 1.045 MESS Adjust OFFSET pot R69 for 0 +/-5mV. 1.046 34401 0mV S 2W 1.047 5520 200degC _E TC S 2W 1.048 MESS Adjust R21 for -1.9600 +/-0.0005V 1.049 34401 -1.9600V S 2W 1.050 JMP 1.061 1.051 LABEL TYPE_T 1.052 HEAD TYPE T THERMOCOUPLE 1.053 DISP Connect the 5520A TC Terminals to the PM-10 input using 1.053 DISP TYPE T thermocouple wire and connectors. 1.054 RNG 200degC 1.055 5520 0degC _T TC S 2W 1.056 MESS Adjust OFFSET pot R69 for 0 +/-5mV. 1.057 34401 0mV S 2W 1.058 5520 200degC _T TC S 2W 1.059 MESS Adjust R21 for -1.9800 +/-0.0005V 1.060 34401 -1.9800V S 2W 1.061 END